Overview

Stereotactic body radiation therapy (SBRT, also called SABR) delivers high doses in a small number of fractions to small, well-defined targets. For lung, that puts particular weight on three things: accounting for respiratory motion, shaping dose tightly around the target with a steep falloff, and respecting nearby organs — which ones depend heavily on whether the tumor is peripheral or central.

This page organizes the concepts and resources behind each step, from simulation to plan evaluation. It explains why each step matters; the numbers — margins, prescriptions, constraints — belong to the protocol you’re following.

Start here: I’m planning my first lung SBRT case — a step-by-step learning roadmap.

Simulation & Motion Management

Lung tumors move with breathing. Everything downstream — target volumes, margins, delivery — depends on how that motion is measured and managed. AAPM TG-76 is the standard reference for this section.

4DCT
A breathing-sorted CT that shows how far, and in which direction, the tumor moves.

What is it?

A CT acquired across the breathing cycle and sorted into phases, so the tumor's position can be viewed at each point of the cycle.

Why does it matter?

A single free-breathing CT is a snapshot: it can catch the tumor anywhere in its path and even distort its shape. 4DCT shows how far and in which direction the tumor actually travels, which drives the choice of motion-management strategy and the target volumes built from it.

Review every phase, not just the derived images. Irregular breathing during the scan can create artifacts that look like changes in tumor shape. Derived datasets — maximum intensity projection (MIP), average intensity projection (AIP) and individual phases — each have a job; which one is used for contouring and which for dose calculation is set by your protocol and physics team.

Respiratory motion
Breathing moves the tumor, mostly up and down, by an amount that varies between patients and tumor locations.

What is it?

Movement of the tumor and surrounding anatomy caused by breathing, usually largest in the superior–inferior direction.

Why does it matter?

Motion varies a lot between patients and between tumor locations; tumors in the lower lobes near the diaphragm tend to move more than apical tumors. Motion that isn't accounted for becomes a geometric miss; motion that is over-accounted for irradiates extra lung.

TG-76 recommends measuring tumor motion for each patient when significant motion is expected — see the report for its thresholds. Breathing at simulation may not match breathing on treatment days; image guidance at treatment is how that difference is caught.

ITV concept
A target volume that covers the tumor's whole path over the breathing cycle.

What is it?

A target volume that encompasses the tumor's positions over the breathing cycle. Concept page →

Why does it matter?

The ITV is the most common way to account for motion in lung SBRT without gating or breath hold: rather than chasing the tumor, the plan covers its whole path.

An ITV is commonly built from the GTV on every phase, or from a MIP. A MIP can under-represent the tumor where it moves next to other high-density tissue (diaphragm, chest wall, mediastinum), so check it against the individual phases.

Gating
Treating in only part of the breathing cycle reduces the motion the plan has to cover.

What is it?

Delivering the beam only during a chosen portion of the breathing cycle.

Why does it matter?

Treating in a narrow window reduces the motion the plan has to cover, which can shrink the target volume for tumors that move a lot.

Gating depends on a breathing signal (an external surrogate or internal markers) that has to correlate with where the tumor really is. Treatment takes longer, and the simulation needs to be acquired and reconstructed for the same gating window.

Breath hold
A reproducible breath hold nearly freezes tumor motion, for patients who can manage it.

What is it?

Treating while the patient holds their breath at a reproducible level.

Why does it matter?

Holding the breath nearly freezes tumor motion for the length of the hold.

Works only for patients who can hold reproducibly — often a challenge for patients with poor lung function. Coaching and a monitoring system help; image guidance verifies the tumor position during the hold.

Abdominal compression
Pressure on the upper abdomen limits diaphragm motion, so a smaller ITV can cover the tumor.

What is it?

Applying pressure to the upper abdomen to limit diaphragm excursion (a “forced shallow breathing” technique in TG-76 terms).

Why does it matter?

It can reduce the amplitude of motion, especially for tumors near the diaphragm, so a smaller ITV covers the tumor's path.

The device position and pressure need to be recorded and reproduced exactly each day, since they change the anatomy and the motion. Patient comfort matters: SBRT sessions can be long.

Image fusion — PET/CT
The staging PET/CT, fused with the planning CT, helps separate tumor from collapsed or scarred lung.

What is it?

Image fusion aligns an outside scan with the planning CT so the target can be drawn using information the CT doesn't show. For lung SBRT this is usually the staging FDG-PET/CT; MRI can also be used. The GTV is still drawn mainly on the planning CT — in lung windows, checked in soft-tissue windows.

Why does it matter?

Tumor can sit against atelectasis or fibrosis that looks much the same on CT. PET helps tell them apart, and one template says it may be necessary when there is a lot of either; IV contrast on the planning CT also helps, because collapsed lung does not enhance. The PET/CT is a separate scan, so check the registration around the tumor — and if a fused image becomes the main image for the GTV, the template suggests an extra margin for registration uncertainty.

Immobilization
Tight margins make reproducible, comfortable positioning more important than in conventional treatment.

What is it?

Devices and positioning that keep setup reproducible across treatment — commonly arms raised above the head for beam access, with an indexed body cushion or frame.

Why does it matter?

SBRT uses tight margins and steep dose falloff, so small setup differences matter more than in conventional treatment.

A position the patient can hold comfortably for the whole session is more reproducible than an ideal position they can't. Image guidance before each fraction checks setup; the immobilization keeps it stable during delivery.

Target Delineation

How target volumes build on one another in lung SBRT. Margins and expansion rules come only from a protocol or institutional policy.

GTV — Gross Tumor Volume

Definition

The visible, palpable or demonstrable extent of tumor.

Why it matters

Every other volume is built from it. In lung, the appearance of the tumor depends on the CT window and level used, so protocols often specify the window for delineation. PET is frequently used alongside CT to distinguish tumor from collapsed lung.

Contouring resources

Organ contouring guides →

ITV — Internal Target Volume

Definition

A volume accounting for internal target motion — in lung, typically derived from 4DCT. Concept page →

Why it matters

Represents the space the tumor occupies over the breathing cycle, so the plan covers the tumor wherever it is during delivery.

Contouring resources

Organ contouring guides →

CTV — Clinical Target Volume

Definition

The volume containing gross tumor plus suspected microscopic disease. Whether a separate CTV expansion is used in lung SBRT depends on the protocol.

Why it matters

Many SBRT protocols rely on the high dose around the tumor edge to treat microscopic extension rather than a separate CTV; others add one. Check which approach your protocol takes before building volumes.

Contouring resources

Organ contouring guides →

PTV — Planning Target Volume

Definition

A geometric expansion that accounts for setup and delivery uncertainty so the intended dose reaches the target.

Why it matters

In SBRT the PTV is also the volume most plan-quality metrics are measured against — coverage, conformity, R50% and D2cm are all defined relative to it — so its size directly shapes how a plan is judged.

Contouring resources

Organ contouring guides →

Central vs Peripheral Tumors

Where the tumor sits changes which structures limit the plan.

  • Peripheral tumors sit out in the lung, often touching the chest wall. The chest wall/ribs, skin and total lung are the usual concerns, and single-fraction or few-fraction schedules have been studied here (see RTOG 0915).
  • Central tumors sit near the mediastinum — the trachea and proximal bronchial tree, esophagus, heart and great vessels. Early experience linked high dose here to serious toxicity, which is why many protocols handle central tumors separately and why RTOG 0813 was designed for them.

Protocols define “central” using distance from the proximal bronchial tree and other mediastinal structures, and the definitions differ. Some groups also describe an “ultracentral” category. Before planning, confirm which definition your protocol uses and which structures it asks you to contour.

Dose Calculation & Delivery

Lung is one of the hardest places to calculate dose accurately: a small, dense tumor surrounded by low-density tissue.

Heterogeneity correction
Lung is far less dense than water, so dose is calculated with the density differences accounted for.

What is it?

Accounting for the different densities of tissue (lung, soft tissue, bone) when calculating dose.

Why does it matter?

Lung tissue is far less dense than water. Ignoring that difference misrepresents the dose actually delivered to the tumor and surrounding lung.

Heterogeneity-corrected calculation is the norm for lung SBRT; AAPM TG-101 discusses it. Your protocol will state what is required.

Calculation algorithm
Algorithms differ in how well they model dose in low-density lung, especially at the edges of small targets.

What is it?

The method the planning system uses to compute dose — for example pencil-beam, convolution/superposition (e.g. AAA, Collapsed Cone), Acuros XB or Monte Carlo.

Why does it matter?

In low-density lung, lateral electron transport increases and the dose at the tumor edge changes. Simpler algorithms model this poorly and can overestimate dose to small targets surrounded by lung.

Know which algorithm your plan was calculated with and whether the protocol restricts it. When comparing plans or reading outcomes data, note the algorithm used — numbers from different algorithms aren't directly comparable.

Beam arrangement & technique
Many beam directions produce the steep dose fall-off that R50% and D2cm measure.

What is it?

SBRT plans typically use many beam directions — multiple static fields, non-coplanar beams or VMAT arcs — to spread entrance dose and concentrate dose at the target.

Why does it matter?

Many beam angles produce the steep dose falloff that metrics like R50% and D2cm measure. Techniques and angle choices also affect skin and chest-wall dose for peripheral tumors.

Collision checks and couch clearance matter for non-coplanar beams. With VMAT and IMRT, small targets with a moving tumor raise the interplay question: how modulated delivery and tumor motion interact. Your physics team's policy governs how much modulation is acceptable.

Image guidance
Imaging before each fraction confirms the tumor sits where the plan expects, which tight margins depend on.

What is it?

Imaging before (and sometimes during) each fraction to confirm the target's position — most often cone-beam CT.

Why does it matter?

With tight margins, daily verification that the tumor sits where the plan expects is central to safe SBRT delivery.

Matching can be on the tumor itself or on bony anatomy; know which your department uses and why. A CBCT of a moving tumor shows a motion-blurred tumor, and its extent should be consistent with the ITV drawn at simulation.

Organs at Risk & Dose Constraints

Click an organ for its anatomy, why it matters and contouring resources. Superscripts link to the study behind each value.

Organ30–34 Gy / 1 fxPeripheral54 Gy / 3 fxPeripheral50 Gy / 5 fxCentral
Spinal Cord
  • D0.35 cc <10 Gy1
  • Max dose <12 Gy2
  • Max dose <18 Gy3
  • D0.25 cc <22.5 Gy4
  • D0.5 cc <13.5 Gy4
Esophagus
  • Max dose <15.4 Gy2
  • Max dose <27 Gy3
  • Max dose 52.5 Gy4
  • D5 cc <27.5 Gy4
Heart / Pericardium
  • Max dose <22 Gy2
  • D15 cc <16 Gy2
  • Max dose <30 Gy3
  • Max dose 52.5 Gy4
  • D15 cc ≤32 Gy4
Great Vessels
  • Max dose <30 Gy2
  • D0.5 cc <45 Gy5
  • Max dose 52.5 Gy4
  • D10 cc <47 Gy4
Trachea & Proximal Bronchial Tree
  • Max dose <20.2 Gy1
  • D4 cc <10.5 Gy1
  • D0.5 cc <30 Gy (optimal), <32 Gy (mandatory)5
  • Max dose 35–38 Gy6
Brachial Plexus
  • Max dose <15 Gy2
  • Max dose <24 Gy3
  • Max dose 32 Gy4
  • D3 cc <30 Gy4
Chest Wall / RibsNever at the expense of PTV coverage.
  • Avoid the 26 Gy isodose through the full thickness of the chest wall2
  • D30 cc <30 Gy (D70 cc <30 Gy acceptable if not achievable)2
  • D30 cc <30 Gy7
Skin
  • Max dose 26 Gy2
  • D10 cc <23 Gy1
  • Max dose 33 Gy6
  • D10 cc <30 Gy6
  • Max dose 32 Gy4
  • D10 cc <30 Gy4
LungsBoth lungs minus iGTV.
  • V20 <10%6
  • Mean ≤8 Gy (mandatory), ≤6 Gy (optimal)6
  • V20 <10%6
  • Mean ≤8 Gy (mandatory), ≤6 Gy (optimal)6
  • V20 <10%6
  • Mean ≤8 Gy (mandatory), ≤6 Gy (optimal)6
  • “Max dose 52.5 Gy” in the five-fraction column equals 105% of the 50 Gy prescription.
  • Spinal cord: bony spinal canal, at least 10 mm above and below the PTV.
References (7)
  1. RTOG 0915 — A Randomized Phase 2 Study Comparing 2 Stereotactic Body Radiation Therapy Schedules for Medically Inoperable Patients With Stage I Peripheral Non-Small Cell Lung Cancer: NRG Oncology RTOG 0915 (NCCTG N0927). Int J Radiat Oncol Biol Phys. 2015;93(4):757–764 Source
  2. TROG SAFRON II — A randomised phase II trial of Stereotactic Ablative Fractionated radiotherapy versus Radiosurgery for Oligometastatic Neoplasia to the lung (TROG 13.01 SAFRON II). BMC Cancer. 2016;16:183 Source
  3. TROG CHISEL — Stereotactic ablative radiotherapy versus standard radiotherapy in stage 1 non-small-cell lung cancer (TROG 09.02 CHISEL): a phase 3, open-label, randomised controlled trial. Lancet Oncol. 2019;20(4):494–503 Source
  4. RTOG 0813 — Safety and Efficacy of a Five-Fraction Stereotactic Body Radiotherapy Schedule for Centrally Located Non–Small-Cell Lung Cancer: NRG Oncology/RTOG 0813 Trial. J Clin Oncol. 2019;37(15):1316–1325 Source
  5. UK SABR Consortium 2019 — Stereotactic Ablative Body Radiation Therapy (SABR): A Resource — UK SABR Consortium guidelines v6.1.0. UK SABR Consortium guidelines, v6.1.0 Source
  6. UK SABR consensus 2022 — UK 2022 Consensus on Normal Tissue Dose-Volume Constraints for Oligometastatic, Primary Lung and Hepatocellular Carcinoma Stereotactic Ablative Radiotherapy. Clin Oncol (R Coll Radiol). 2022;34(5):288–300 Source
  7. Lo 2013 — Serious complications associated with stereotactic ablative radiotherapy and strategies to mitigate the risk. Clin Oncol (R Coll Radiol). 2013;25(6):378–387 Source

Values quoted from eviQ 3389 (last reviewed 16 March 2023; review due 30 June 2027) and eviQ 4018 (last reviewed 16 March 2023; review due 30 June 2027), checked by DoseLab on 2026-09-22. Verify against your protocol, institutional policy and physician directive before clinical use.

Search all lung constraints in the Constraint Finder →

Protocols & Literature

Trials and reports every lung SBRT planner should know exist. Bibliographic details marked Under Review are awaiting DoseLab's check against the original.

Landmark trialUnder Review

Randomized phase 2 comparison of a single-fraction and a multi-fraction SBRT schedule for peripheral stage I non-small cell lung cancer. An erratum and a long-term follow-up report were also published.

Consensus guidelineSource Verified

UK consensus on dose–volume constraints for SABR of oligometastases, primary lung cancer and hepatocellular carcinoma. Citation checked against Crossref. DoseLab read the tables in the authors' accepted manuscript (UCL Discovery and White Rose Research Online): Table 6 (intracranial, 1, 3 and 5 fractions) for the brain pages, and Tables 2, 3 and 5 (thoracic, abdominal and spinal) for the liver SBRT page. Each table gives optimal and mandatory limits by fraction count; the liver limits apply to patients with, at worst, Child-Pugh A6 liver disease. The kidney limit is written as a "cold" constraint — at least 200 cc must receive no more than the stated dose.

AAPM reportSource Verified

AAPM task group report on respiratory motion in thoracic, abdominal and pelvic treatment. It groups management approaches into motion-encompassing methods, respiratory gating, breath-hold techniques, forced shallow breathing (such as abdominal compression) and respiration-synchronized (tracking) techniques. Read in the AAPM report PDF (Report No. 91, July 2006): using ICRU Report 62 nomenclature, the PTV adds margins to the CTV for respiratory (intrafraction) motion, interfraction motion and set-up error; without a respiratory management device the entire range of motion should be considered in the internal margin; and for lung tumors a maximum intensity projection (MIP) can give the motion-encompassing volume, provided there is no mediastinal involvement.

Review articleSource Verified
A Story of Hypofractionation and the Table on the Wall
Timmerman R · UT Southwestern Medical Center · 2022

An editorial history of the UT Southwestern SABR constraint tables — often called the “Timmerman tables” — covering how they were developed and revised, and their relationship to AAPM TG-101 and NRG/RTOG protocols. Tables 2–11 separate serial tissues (listed with a volume, a volume max and a max point dose) from parallel tissues (listed with a critical volume and a critical-volume dose max). In Tables 2–7 a footnote defines “point” as ≤0.035 cm³; Tables 8–11 add a contouring-instructions column for each structure.

All lung literature →